European Ocean Biodiversity Information System

[ report an error in this record ]basket (0): add | show Print this page

Solid microelectrodes for in situ voltammetric measurements
Pizeta, I.; Billon, G.; Fischer, J.-C.; Wartel, M. (2003). Solid microelectrodes for in situ voltammetric measurements. Electroanalysis 15(17): 1389-1396. https://dx.doi.org/10.1002/elan.200302669
In: Electroanalysis. Wiley-VCH: Weinheim. ISSN 1040-0397; e-ISSN 1521-4109
Peer reviewed article  

Available in  Authors 
    Vlaams Instituut voor de Zee: Open access 61790 [ download pdf ]

Keyword
    Marine/Coastal

Authors  Top 
  • Pizeta, I.
  • Billon, G.
  • Fischer, J.-C.
  • Wartel, M.

Abstract
    Solid Au and Ag microelectrodes for both,model,and in situ voltammetric measurements, have been designed andtested. Amalgamation procedure and the parameters of measurement procedure are discussed. Differences betweengold and silver electrodes were studied. Silver electrodes showed more stable behavior and were used prevalently for calibration and in situ measurements.Calibration of a 30-µm diameter Ag electrode for Mn(II), Fe(II), I(-I) andS(-II) showed good linearity with reasonable confidence intervals, necessary for quantification of the signalsobtained in situ, with average sensitivity of 0.05, 0.01, 0.16 and 1.10 nA/(µmol L-1) and minimal detection limit of 3.5, 28, 0.3 and 0.05 µmol L-1, respectively. Additional checking of the electrode sensitivity with the reduction peaks of oxygen is proposed.

All data in the Integrated Marine Information System (IMIS) is subject to the VLIZ privacy policy Top | Authors